Analysis of Wedge Angle Variations in the Treatment Planning System Based on Dose Volume Histogram on Ca Mammae Sinistra

  • Sri Nurjannah Mulawarman University
  • Robert Janssen Stevenly A. W. Sjahranie Hospital
  • Kadek Subagiada Mulawarman University
  • Erlinda Ratnasari Putri Mulawarman University
Keywords: Breast Cancer, Organs at Risk, Planning Target Volume, Radiotherapy, Wedge

Abstract

Breast cancer ranks first in terms of the highest number of cancers in Indonesia and is one of the first contributors to cancer deaths. According to IARC, radiotherapy is one of the most important modalities in the treatment of breast cancer. Radiotherapy is a type of treatment that uses ionizing radiation to kill cancer cells without damaging surrounding healthy tissue. To obtain an even or homogeneous dose distribution on an uneven skin surface, a wedge is used so that the radiation given to the tumor area is truly optimal and the surrounding healthy tissue receives a minimum dose. This research was carried out by evaluating the Planning Target Volume (PTV) value based on ICRU-83 and dose distribution in Organs at Risk (OAR) based on QUANTEC standards at four different wedge angles, namely 15°, 30°, 45°, and 60°, used the 3D-CRT technique of tangential radiation with a dose of 50 Gy in 25 fractions on 5 left-sided breast cancer patients. The research results showed that the average Homogeneity Index (HI) values for the 15°, 30°, 45° and 60° wedges were 0.286, 0.346, 0.436 and 0.578, respectively. In the left and right lung dose distribution, the results obtained for the 15°, 30°, 45°, and 60° wedge were respectively 2231.4 and 116.2 cGy, 2257.8 and 114.6 cGy, 2141 and 100 cGy, and 2059.4 and 88.8 cGy. For the average dose distribution to the heart, the values obtained at the 15°, 30°, 45°, and 60° wedges were 2005.6 cGy, 2040.4 cGy, 1930.6 cGy, and 1844.4 cGy, respectively. where the average dose received by OAR is still below the limit allowed by QUANTEC. Using a wedge angle of 15° provides a homogeneous dose distribution for the PTV and a wedge angle of 60° provides the lowest dose distribution for the OAR.

Downloads

Download data is not yet available.

References

Rizka, Adi, Muhammad Khalilul Akbar, and Narisha Amelia Putri. 2022. “Carcinoma Mammae Sinistra T4bN2M1 Metastasis Pleura.” AVERROUS: Jurnal Kedokteran dan Kesehatan Malikussaleh 8(1): 23.

Kania, Nia. 2018. Payudara Dan Kelainannya. Banjarbaru: PT. Grafika Wangi Kalimantan. http://eprints.ulm.ac.id/3851/1/Buku Nia Kania Upload Repository ULM.pdf.

American Cancer Society. 2022. “Breast Cancer Facts & Figures 2022-2024.” American Cancer Society, Inc.

Susworo, R., and Kodrat Henry. 2017. Radioterapi : Dasar-Dasar Radioterapi Tata Laksana Radioterapi Penyakit Kanker. Jakarta: UI-Press.

N. Alya, R. J. Stevenly, P. S. Wardani, and E. R. Putri, “Effect of Radiation Position of Photon Beam at Axis Point and Field Edge on Absorbed Dose on Linear Accelerator (Linac) ”, JI, vol. 9, no. 1, pp. 103-110, Mar. 2024.

E. B. Podgorsak, “New mannose-binding lectin from the rhizome of sarsaparilla Smilax glabra Roxb. (Liliaceae),” J. Agric. Food Chem., vol. 52, no. 20, pp. 6091–6095, 2005, doi: 10.1021/jf030837o.

Khan, Faiz M. 1994. The Physics of Radiation Therapy. Second Edi. Universitas Michigan: Williams & Wilkins.

Idris, Rilwanu, and Hassan Ibrahim. 2019. “Dosimetric Consequences of Using Wedge Angles as Major Determinants of Tumor Dose Homogeneity in Breast Cancer Teletherapy.” African Journal of Medical Physics 2(2).

Fitriyah, Nurul, Rahmatul Izza Nur Amalia, Bambang Haris Suhartono, and Suryani Dyah Astuti. 2020. “The Effect of Dynamic Wedge Angle in Breast Cancer’s Absorbed Radiation Dose.” 2314.

Saukah, Hulfi Azmy. 2022. “Analisis Variasi Wedge Pada Treatment Planning System Teknik 3D-CRT Terhadap Dose Volume Histogram Untuk Kasus Kanker Payudara.” Universitas Nasional Jakarta.

Husni, Mutiatul, Mohammad Ali Shafii, Rico Adrial, and Muhammad Ilyas. 2021. “Analisis Perbandingan Nilai Conformity Index Dan Homogeneity Index Pada Teknik 3D-CRT Dan IMRT Pada Kasus Kanker Payudara Berdasarkan Hasil TPS Di RS UNAND.” Jurnal Fisika Unand 10(4): 511–17.

Cao, Tingting et al. 2019. “Analysis of Different Evaluation Indexes for Prostate Stereotactic Body Radiation Therapy Plans: Conformity Index, Homogeneity Index and Gradient Index.” Precision Radiation Oncology 3(3): 72–79.

Published
2024-09-09
How to Cite
[1]
S. Nurjannah, Robert Janssen Stevenly, Kadek Subagiada, and Erlinda Ratnasari Putri, “Analysis of Wedge Angle Variations in the Treatment Planning System Based on Dose Volume Histogram on Ca Mammae Sinistra”, JI, vol. 9, no. 2, pp. 420-427, Sep. 2024.